Anton Paar Alcohol Analysis System: Alcolyzer Plus Spirits
Analyze spirits efficiently and economically in the laboratory and at the production line using the alcohol analysis system Alcolyzer Plus Spirits. Alcolyzer Plus Spirits determines the alcohol content (35 to 65 %vol), color, and pH value of alcoholic beverages. In combination with a density meter, the system additionally determines the total extract of the sample.
Alcohol analysis system with patented NIR measuring method
- Highly accurate measurement of the alcohol content of your spirits in the concentration range 35 to 65 %vol
- Optional pH and color measurement
- For spirits such as whisky, cognac, brandy, vodka, gin, tequila and rum
Time-saving and user-friendly
- Ready to measure after only 10 minutes
- Only 3 to 4 minutes from filling to the result
- Accurate, fully automatic thermostatting via an integrated solid-state thermostat
Rapid and easy adjustment
- Adjustment/calibration with water and an alcohol/water solution
- One single adjustment is valid for all your samples
Use at the production line or in the lab
- Integrated peristaltic pump for use at the production line
- Combine with a density meter and sample changer for lab measurements
Industries
- Beverages
Specifications
Whisky | For spirits with an extract content up to 5 g/L (e.g. whisky, vodka, gin, tequila and rum) | |||
---|---|---|---|---|
Cognac | For spirits with an extract content up to 20 g/L (e.g. cognac and brandy) | |||
Measuring range | ||||
Alcohol | 35 to 65 %vol (data can be displayed from 0 to 90 %vol) | |||
pH value (optional) | 0 to 14 | |||
Color (optional) | 0 to 120 EBC | |||
Density (optional) | 0 to 3 g/cm3 | |||
Repeatability - standard deviation | ||||
Alcohol | ||||
Manual filling | 0.03 %vol | |||
Automatic filling | 0.01 %vol | |||
pH value (optional) | 0.02 | |||
Color (optional) | 0.1 EBC | |||
Density (optional) | 0.00001 g/cm3 (DMA 4500) or 0.000001 g/cm3 (DMA 5000) | |||
Additional information | ||||
Interfaces | 2 x RS-232 for printer, PC and DMA | |||
Connection | For IBM-compatible keyboard and/or bar code reader |
Lab system | Production system | ||
Sample volume | 30 mL per measurement | 30 mL per measurement | |
Typical measuring duration per sample | 4 minutes (incl. filling) | 3 minutes (incl. filling) | |
Sample throughput | 15 samples per hour | 20 samples per hour | |
Weight | approx. 53.4 kg (117.7 lbs) | approx. 12.4 kg (26 lbs) | |
Dimensions (L x W x H) | 590 x 690 x 530 mm (23.2 x 27.2 x 20.9 inches) |